Masters theses : Science

Permanent URI for this collectionhttps://repository.neelain.edu.sd/handle/123456789/12103

Browse

Search Results

Now showing 1 - 1 of 1
  • Thumbnail Image
    Item
    A METHOD FOR ASSESSING GAMMA CAMERA LESION DETECTABILITY
    (ALNEELAIN UNIVERSITY, 1999-11) GAF AR HAYDER MOHAMMED SAEED
    its/<§§_§_l'//,"iA.é/_¢/T A new simple method for the assessment of the capability of a gamma camera in detecting lesions within the patient is described. The method, when applied on the ZLC-37 Siemens gamma camera modefied by Mediso, showed that this gamma camera equiped with a ( LEAP) collimator was capable of detecting lesions of dimensions of; 0.4 cm,0.6 cm, 0.8 cm, 0.9 cm, 1.0 cm and 1.2 cm and above, when they were located at; the surface of the patient, depth of 4 cm or less, depth of 9 cm or less, depth of 1 1 cm or less, depth of l4 cm or less and l9 cm or less, respectively. When these values were compared with similar results obtained from a thyroid phantom studies, showed good agreement. Further studies to develop the method were recornmended. 1 l mil