A METHOD FOR ASSESSING GAMMA CAMERA LESION DETECTABILITY

Thumbnail Image

Date

1999-11

Journal Title

Journal ISSN

Volume Title

Publisher

ALNEELAIN UNIVERSITY

Abstract

its/<§§_§_l'//,"iA.é/_¢/T A new simple method for the assessment of the capability of a gamma camera in detecting lesions within the patient is described. The method, when applied on the ZLC-37 Siemens gamma camera modefied by Mediso, showed that this gamma camera equiped with a ( LEAP) collimator was capable of detecting lesions of dimensions of; 0.4 cm,0.6 cm, 0.8 cm, 0.9 cm, 1.0 cm and 1.2 cm and above, when they were located at; the surface of the patient, depth of 4 cm or less, depth of 9 cm or less, depth of 1 1 cm or less, depth of l4 cm or less and l9 cm or less, respectively. When these values were compared with similar results obtained from a thyroid phantom studies, showed good agreement. Further studies to develop the method were recornmended. 1 l mil

Description

A THESIS SUBMETTED IN F ULF ILMENT FOR THE DEGREE OF MASTER OF SCINCE IN PHYSICS - SUDAN UNIVERSITY OF SCEINCE & TECHNOLOGY.

Keywords